Operator’s Manual  ZM-520PA/521PA/530PA/531PA
SpO2 RESPONSE
Select the SpO
2
 response mode.
FAST: 
Select this for special applications that require a fast response. FAST is suitable for  
 
detecting short apnea.
NORMAL:  For normal monitoring.
SLOW: 
Select this when you need to suppress a rapid change in SpO
2
.
1.  On the PARAMETER SETUP screen, press 
the 
▼ key to move the cursor to “SpO2 
RESPONSE”. “SpO2 RESPONSE” is on the 
second page of the PARAMETER SETUP 
screen.
2.  Press the Function key.
3. Press 
the 
▼ key to select “FAST”, 
“NORMAL” or “SLOW”.
4.  Press the Function key to register the 
selected setting. The cursor returns to “SpO2 
RESPONSE”.
PARAMETER SETUP screen - page 2
PARAMETER SETUP screen - page 2