Operator’s Manual  ZM-520PA/521PA/530PA/531PA
 31
Changing SYSTEM SETUP Settings
System Setup Setting List
The factory default settings are underlined.
Setting Item
Description
Settings
PROTOCOL
Select the transmitting protocol.
51: New protocol. A central monitor with ORG-
9100A/9110A multiple patient receiver whose 
software version 03-03 or later can receive this 
protocol.
41: Old protocol. A central monitor with ORG-
9100A/9110A/9700A multiple patient receiver can 
receive this protocol.
NOTE
When 51 is set, the receiving monitor must be able 
to receive protocol 51. Otherwise, signals from the 
transmitter cannot be received. To use protocol 
51, the ORG-9100A/ORG-9110A multiple patient 
receiver software version 03-03 or later is required.
51, 41
BRIGHTNESS
Select the screen brightness.
DARK, 
BRIGHT
FUNCTION KEY
Select the function of the Function key.
SUSPEND ALARM & PAUSE:
 
Suspends alarm on the receiving monitor for 2 
minutes. Pauses monitoring on the transmitter 
and receiving monitor.
SUSPEND ALARM:
 
Suspends alarm on the receiving monitor for 2 
minutes.
CONFIRM:
 
Displays the “PATIENT CONFIRMED” message 
on the transmitter screen and transmits the 
message to the receiving monitor.
OFF: No 
function.
NOTE
“SUSPEND ALARM & PAUSE” and “CONFIRM” can 
only be set when PROTOCOL is set to 51. To use 
protocol 51, the ORG-9100A/ORG-9110A multiple 
patient receiver software version 03-03 or later is 
required.
SUSPEND 
ALARM 
& PAUSE, 
SUSPEND 
ALARM, 
CONFIRM, 
OFF