Operator’s Manual  ZM-520PA/521PA/530PA/531PA
NOTE
FUNCTION KEY (on the second page of 
the SYSTEM SETUP screen) can be set to 
“SUSPEND ALARM & PAUSE” or “CONFIRM” 
only when PROTOCOL is “51”. If PROTOCOL 
is changed to “41”, FUNCTION KEY is 
automatically changed to “OFF”.
4.  Press the Function key to register the selected 
setting. The cursor returns to “PROTOCOL”.
BRIGHTNESS
Select the screen brightness.
1. Press 
the 
▼ key to move the cursor to “BRIGHTNESS”.
2.  Press the Function key.
3. Press 
the 
▼ key to select “DARK” or “BRIGHT”.
4.  Press the Function key to register the selected setting. The cursor returns to 
“BRIGHTNESS”.
FUNCTION KEY
Select the function of the Function key. For details on using these functions, refer to “Basic 
Monitoring Operation” in the “Monitoring” section.
SUSPEND ALARM & PAUSE:  Suspends alarm on the receiving monitor for 2 minutes. Pauses  
 
monitoring on the transmitter and receiving monitor.
SUSPEND ALARM: 
Suspends alarm on the receiving monitor for 2 minutes.
CONFIRM: 
Displays the “PATIENT CONFIRMED” message on the  
 
transmitter screen and transmits the message to the receiving  
 monitor.
OFF: No 
function.
NOTE
“SUSPEND ALARM & PAUSE” and “CONFIRM” can only be set when PROTOCOL is set 
to 51.