Operator’s Manual  ZM-520PA/521PA/530PA/531PA
 33
Changing System Setup Settings
PROTOCOL
Select the transmitting protocol. For differences between protocols, refer to the table below.
51:  New protocol. A central monitor with ORG-9100A/9110A multiple patient receiver    
 
whose software version 03-03 or later can receive this protocol.
41:  Old protocol. A central monitor with ORG-9100A/9110A/9700A multiple patient receiver  
 
can receive this protocol.
NOTE
When 51 is set, the receiving monitor must be able to receive protocol 51. Otherwise, 
signals from the transmitter cannot be received.
Differences Between Protocols
Function
Protocol 41
Protocol 51
Setting ECG MEASUREMENT to OFF on 
the transmitter automatically turns off the 
ECG measurement setting on the receiving 
monitor 
No (ECG measurement 
must be turned off on 
the receiving monitor)
Yes
Pause monitoring on the receiving monitor 
from the transmitter
No
Yes
Transmit “PATIENT CONFIRMED” message No
Yes
Display battery level of the transmitter on the 
receiving monitor
No
Yes
Transmit SpO
2
 messages 
Some messages (refer 
to the “Indication and 
Message List” section)
All messages
Transmit RESP message
No
Yes
1. Press 
the 
▼ key to move the cursor to 
“PROTOCOL”.
2.  Press the Function key.
3. Press 
the 
▼ key to select “51” or “41”.
Setting
Cursor
Setting
Cursor