Operator’s Manual  ZM-520PA/521PA/530PA/531PA
Suspending Alarms on the Receiving Monitor
When the FUNCTION KEY is set to “SUSPEND ALARM” or “SUSPEND ALARM & PAUSE” 
on the SYSTEM SETUP screen, alarms can be suspended for 2 minutes on the receiving monitor 
before they occur.
To suspend alarms:
1.  Press the Function key. The “Suspend alarms” confirmation screen appears.
2. Press 
the 
▼ key to suspend alarms.
 
To cancel suspending alarms and return to the previous screen, press the Screen key.
When the alarms are suspended, the “ALARMS SUSPENDED” message and alarm suspended 
icon with the remaining minutes in alarm suspension appear on the transmitter screen.