Operator’s Manual  ZM-520PA/521PA/530PA/531PA
Panel Description
Top Panel
For attaching a strap.
ECG/RESP socket
Connects the electrode lead for 
measuring ECG and/or respiration by 
the impedance method.
ZM-530PA/531PA only
SpO
2
 socket
Connects the SpO
2
 probe.
WARNING
Before defibrillation, all persons must 
keep clear of the bed and must not 
touch the patient or any equipment or 
cord connected to the patient. Failure 
to follow this warning may cause 
electrical shock or injury.
WARNING
When performing defibrillation, 
discharge as far as possible from 
electrodes, patches and any gel, 
cream or medicine on the chest of the 
patient. If there is a possibility that the 
defibrillator paddle could touch these 
materials, remove them from the 
patient. If the defibrillator paddle 
directly contacts these materials, the 
discharged energy may cause skin 
burn to the patient.
WARNING
When the transmitter is used with an 
electrosurgical unit (ESU), firmly attach 
the entire area of the ESU return plate. 
Otherwise, the current from the ESU 
flows into the electrodes of the 
transmitter, causing electrical burn 
where the electrodes are attached. For 
details, refer to the ESU manual.
CAUTION
Do not shake or swing the transmitter 
while holding the leads or cables 
connected to the transmitter. The 
transmitter may come off and injure 
someone or damage surrounding 
instruments.