Operator’s Manual ZM-520PA/521PA/530PA/531PA
Panel Description
Top Panel
For attaching a strap.
ECG/RESP socket
Connects the electrode lead for
measuring ECG and/or respiration by
the impedance method.
Connects the electrode lead for
measuring ECG and/or respiration by
the impedance method.
ZM-530PA/531PA only
SpO
2
socket
Connects the SpO
2
probe.
WARNING
Before defibrillation, all persons must
keep clear of the bed and must not
touch the patient or any equipment or
cord connected to the patient. Failure
to follow this warning may cause
electrical shock or injury.
keep clear of the bed and must not
touch the patient or any equipment or
cord connected to the patient. Failure
to follow this warning may cause
electrical shock or injury.
WARNING
When performing defibrillation,
discharge as far as possible from
electrodes, patches and any gel,
cream or medicine on the chest of the
patient. If there is a possibility that the
defibrillator paddle could touch these
materials, remove them from the
patient. If the defibrillator paddle
directly contacts these materials, the
discharged energy may cause skin
burn to the patient.
discharge as far as possible from
electrodes, patches and any gel,
cream or medicine on the chest of the
patient. If there is a possibility that the
defibrillator paddle could touch these
materials, remove them from the
patient. If the defibrillator paddle
directly contacts these materials, the
discharged energy may cause skin
burn to the patient.
WARNING
When the transmitter is used with an
electrosurgical unit (ESU), firmly attach
the entire area of the ESU return plate.
Otherwise, the current from the ESU
flows into the electrodes of the
transmitter, causing electrical burn
where the electrodes are attached. For
details, refer to the ESU manual.
electrosurgical unit (ESU), firmly attach
the entire area of the ESU return plate.
Otherwise, the current from the ESU
flows into the electrodes of the
transmitter, causing electrical burn
where the electrodes are attached. For
details, refer to the ESU manual.
CAUTION
Do not shake or swing the transmitter
while holding the leads or cables
connected to the transmitter. The
transmitter may come off and injure
someone or damage surrounding
instruments.
while holding the leads or cables
connected to the transmitter. The
transmitter may come off and injure
someone or damage surrounding
instruments.