Nihon Kohden Corporation ZM-930PA Benutzerhandbuch

Seite von 70
Operator's Manual  ZM-920PA/930PA
1
Introduction
The ZM-920PA/930PA transmitter transmits ECG and other data from a patient to
a Nihon Kohden monitor for continuous monitoring.  Available parameters and
functions vary between the models.  Read the operator’s manual for the monitor
before operation.
Model Parameters 
Functions 
ZM-920PA 
• ECG 
• Impedance respiration 
The following information is 
indicated by LED. 
• Check ECG electrodes 
• Replace batteries 
ZM-930PA 
• ECG 
• Impedance respiration 
• SpO
2
 
The following information is 
indicated on LCD. 
• SpO
2
 value 
• Pulse wave amplitude 
• Replace batteries 
The following information is 
indicated by LED. 
• Check ECG electrodes 
 
The transmitter channel can be changed by the QI-901PK Channel Writer.  To
change the channel number, refer to the channel writer manual.
WARNING
The following actions must be taken to properly receive the
transmitter signal of the correct patient on the receiving instrument,
otherwise, there may be signal loss or signals may mix causing a
serious accident, such as monitoring a different patient.
••••• Assign a channel administrator in the hospital and only he or she
should manage channel assignments.
••••• The channel administrator must manage the channels in the
facility so that there is no signal interference.
••••• When the transmitter channel is changed, the channel
administrator must check that the channel on the receiving
monitor is also changed and the signal is properly received.