Nihon Kohden Corporation ZM-930PA Benutzerhandbuch

Seite von 70
Operator's Manual  ZM-920PA/930PA
7
CAUTION
• Only use Nihon Kohden specified electrodes, electrode leads and
SpO
2
 probes. Otherwise, the maximum performance from the
instrument cannot be guaranteed.
• Do not reuse disposable items.
• Do not shake or swing the transmitter while holding the leads or
cables connected to the transmitter.  The transmitter may come off
and injure someone or damage surrounding instruments.
• Attach a strap to the transmitter to prevent the transmitter from
falling.
• Turn off the power of mobile phones, small wireless devices and
other devices which produce strong electromagnetic interference
around a patient (except for devices allowed by the hospital
administrator).  Radio waves from devices such as mobile phones
or small wireless devices may be mistaken as respiration or pulse
waves and the displayed data may be incorrect.
• Do not use the same channel for different patients.  If the same
channel is used for two patients, the two patients’ data will be lost
due to mutual modulation interference, or another patient’s data
may appear on the receiving monitor screen.
• Do not use transmitters of adjacent channels in a hospital.  If a
transmitter of an adjacent channel is used, radio waves from one
transmitter affects the receiver of the adjacent channel’s
transmitter and there may be interference.
• Close the battery case cover during operation.  If the transmitter is
used with the battery case cover open, anyone who touches the
opened battery case may receive an electrical shock when
defibrillation is performed.  Touching the opened battery case may
cause electrostatic discharge and intermittently interfere with the
waveform or data.