Operator's Manual  ZM-920PA/930PA
Important Safety Information
General
WARNING
• Never use the transmitter in the presence of any flammable
anesthetic gas or high concentration oxygen atmosphere.  Failure
to follow this warning may cause explosion or fire.
• Never use the transmitter in a hyperbaric oxygen chamber.
Failure to follow this warning may cause explosion or fire.
• Never take this transmitter into an MRI test room.
• When performing MRI test, remove all electrodes from the patient
which are connected to this instrument.  Failure to follow this
warning may cause skin burn on the patient.  For details, refer to
the MRI manual.
• Before defibrillation, check that the electrode leads and SpO
2
probe attached to the patient are properly connected to the
transmitter.  Touching the metal parts of disconnected leads and
probes may cause electrical shock or injury by discharged energy.
• When performing defibrillation, discharge as far as possible from
electrodes, patches and any gel, cream or medicine on the chest
of the patient.  If there is a possibility that the defibrillator paddle
could touch these materials, remove them from the patient.  If the
defibrillator paddle directly contacts these materials, the
discharged energy may cause skin burn to the patient.
• Before defibrillation, all persons must keep clear of the bed and
must not touch the patient or any equipment connected to the
patient.  Failure to follow this warning may cause electrical shock
or injury.
• When the transmitter is used with an electrosurgical unit (ESU),
firmly attach the entire area of the ESU return plate.  Otherwise,
the current from the ESU flows into the electrodes of the
transmitter, causing electrical burn where the electrodes are
attached.  For details, refer to the ESU manual.