To open the clip, firmly pull out
the tab in direction of the arrow.
NOTE
••••• When using rechargeable NiCd batteries or NiMH batteries,
shallow charging/discharging shortens battery capacity.  For
details, refer to the battery operator’s manual.
••••• Remove the batteries from the transmitter before disposing of the
transmitter.
Situations Requiring Battery Replacement
Replace the batteries when any of the following occurs.
• The “ ” LED lights (ZM-920PA) or the “ ” mark is displayed on the LCD
(ZM-930PA) on the transmitter.
• The transmitter generates a constant alarm (continuous “peep” sound).
• The monitor displays the battery replacement message on the screen.
• When the power of the LCD transmitter is turned on, no message or icon is
displayed.  (Only the ZM-930PA).
Attaching a Strap to the Transmitter
NOTE
••••• Attach a strap to the transmitter to prevent the transmitter from
falling.
••••• Do not attach the clip to hard objects such as thick cloths or
zippers, or the clip may break.
Attach a strap to the transmitter and fasten the clip to the patients’ clothes or bed
sheets.
If the transmitter falls off, the battery cover may be opened.  If the  patient touches
the terminals of the batteries, patient leakage current over the allowable amount
can occur.