Operator's Manual  ZM-920PA/930PA
WARNING
When the transmitter is used with an electrosurgical unit (ESU),
firmly attach the entire area of the ESU return plate.  Otherwise, the
current from the ESU flows into the electrodes of the transmitter,
causing electrical burn where the electrodes are attached.  For
details, refer to the ESU manual
NOTE
••••• This transmitter is not protected against noise generated from an
electrosurgery unit.
••••• If an electric blanket is used and incorrect heart rate is displayed
on the monitor, turn off the pacing pulse detection on the monitor.
ECG Measurement Procedure
1.
Select the type of electrode lead and disposable electrode according to the
purpose.
2.
Connect the electrode lead to the ECG/RESP socket.
3.
Connect disposable electrodes to the electrode lead and attach electrodes to
the patient.
After steps 1 to 3 are finished, ECG monitoring automatically starts.