Nihon Kohden Corporation ZM-930PA Benutzerhandbuch

Seite von 70
Operator's Manual  ZM-920PA/930PA
23
RA
LA
N (RL)
LL
Va
Vb
Six Electrodes
Electrode Position
The 6-electrode method with lead II and lead V5 is effective for monitoring
myocardial ischemia.  You can improve monitoring accuracy considerably by
adding lead V4 to this combination.  Va and Vb can be at any position of the
standard 12 leads V1 to V6, but V4 and V5 are most appropriate for myocardial
ischemic monitoring.
Symbol 
Lead Color (Clip Color) 
Electrode Position 
RA 
White (White) 
Right infraclavicular fossa 
LA 
Black (Black) 
Left infraclavicular fossa 
LL Red 
(Red) 
Lowest rib on the left anterior 
axillary line 
N (RL) 
Green (Green) 
Right anterior axillary line at the 
same level as LL 
Va 
Brown (Brown) (BR-906PA) 
Brown-blue (BR-916PA) 
Fifth intercostal space on the left 
midclavicular line. (V4 position of 
standard 12 leads) 
Vb 
Brown (Brown) (BR-906PA) 
Brown-orange (BR-916PA) 
Left anterior axillary line at the 
same level as Va. (V5 position of 
standard 12 leads)