Operator's Manual  ZM-920PA/930PA
CAUTION
• Turn off the power of mobile phones, small wireless devices and
other devices which produce strong electromagnetic interference
around a patient (except for devices allowed by the hospital
administrator).  Radio waves from devices such as mobile phones
or small wireless devices may be mistaken as respiration or pulse
waves and the displayed data may be incorrect.
• Normal external light does not affect monitoring but strong light
such as a surgical light or sunlight may affect monitoring.  If
affected, cover the measuring site with a blanket.
• Do not pull or bend the probe cable, and do not put caster feet on
the probe cable.  Do not immerse the probe cable in chemical
solutions or water.  Failure to follow these instructions may cause
cable discontinuity, short circuit, skin burn on the patient and
incorrect measurement data.  Replace any broken probe with a
new one.
• When the probe is attached on an appropriate site with sufficient
circulation and the error message confirming the probe
attachment repeatedly appears, the probe may be deteriorated.
Replace it with a new one.
• While a patient is on medication which causes vasodilation, the
pulse waveform may change and in rare cases the SpO
2
 value
may not be displayed.
NOTE
When monitoring SpO
2
, monitor ECG at the same time.  The ECG
electrode lead works as an antenna for transmitting data from the
transmitter to the receiving monitor.  If ECG is not measured, the
telemetry signal may not be received.