Operator's Manual  ZM-920PA/930PA
Position 3
In this position, the respiration waveform is optimum, but the ECG lead is
unusual.
Position 4
In this position, the respiration measurement is influenced by the impedance
variation of the abdomen, so the cardiac pulse wave included in the respiration
wave is reduced.  Note that the waveform is inverted in phase compared with the
chest movement (the waveform goes down during inspiration).  It is difficult to
measure the ECG at the same time.
Position 2
In this position, the waveform amplitude is usually large and the ECG lead is
similar to Lead MII.  This position can be generally recommended.
RA
 
LL
 
Right infraclavicular 
fossa 
Fifth intercostal space on 
the left midaxillary line, V6
RA
 
LL
 
Right midaxillary at the 
horizontal level of V4 
Fifth intercostal space on the 
left midaxillary line, V6 
RA
 
LL
 
Lowest rib on the right 
anterior axillary line 
Lowest rib on the left 
anterior axillary line