Operator's Manual  ZM-940PA
23
Changing Parameter Setup Settings
The initial settings on the PARAMETER SETUP screen must be changed before monitoring.
Changing these settings during monitoring interrupts monitoring.
Parameter Setup Setting List
The factory default settings are underlined.
Setting Item
Description
Settings
SELECTABLE
INTERVALS
Select the NIBP measurement modes for
the mode selection.
STAT, 5, 10, 15, 30, 60,
120, 240
INITIAL
INTERVAL
Select the initial NIBP measurement
mode at power on.
MAN., 5 min, 10 min,
15 min, 30 min, 60 min,
120 min, 240 min
INITIAL CUFF
PRESS
Select the NIBP cuff inflation pressure.
120 mmHg, 150 mmHg,
180 mmHg, 210 mmHg,
240 mmHg
NIBP MODE
AFTER STAT
Select the NIBP measurement mode
after completing STAT measurement.
MAN., 5 min, 10 min,
15 min, 30 min
START/FINISH
SOUND
Turn ON or OFF the sound for NIBP
measurement start/finish.
ON, OFF/ON, OFF
OLD NIBP DATA
                    AFTER
Select whether to hide or dim the NIBP
data after measurement and how long to
wait after measurement to dim or hide it.
DATA: HIDE, DIM
AFTER: 5 min, 10 min,
30 min
INHIBIT SpO
2
DURING NIBP
Turn SpO
2
 monitoring on or off during
NIBP measurement.
ON, OFF
2ND PARAMETER
Set SpO
2
 and PR display order.
SpO
2
, PR
LEADS OFF
DISPLAY
Select the mode for displaying electrode
off. This setting is only available when
ECG is monitored with 6 electrodes.
CHAR, IMAGE
ECG ELECTRODE
Select the electrode lead type. This
setting is only available when CHAR is
selected for LEADS OFF DISPLAY.
IEC, AHA