Nihon Kohden Corporation ZM-940PA Benutzerhandbuch

Seite von 111
Operator's Manual  ZM-940PA
25
INITIAL INTERVAL
Select the initial NIBP measurement mode at power on.
1.
Press the NIBP INTERVAL key to move the
cursor to “INITIAL INTERVAL”.
2.
Press the NIBP START/STOP key to select
the mode.  Selectable modes are the modes
selected for “SELECTABLE INTERVALS”
and “MAN.” (MANUAL).
INITIAL CUFF PRESS
Select the NIBP cuff inflation pressure.
1.
Press the NIBP INTERVAL key to move the
cursor to “INITIAL CUFF PRESS”.
2.
Press the NIBP START/STOP key to select
the inflation pressure from 120, 150, 180, 210
and 240 mmHg.
1.
Press the NIBP INTERVAL key to move the
cursor to the desired mode.
2.
Press the NIBP START/STOP key to select or
unselect the mode.  Selectable modes are: STAT,
5, 10, 15, 30, 60, 120 and 240 min.
Changing Settings
SELECTABLE INTERVALS
During monitoring, when the NIBP INTERVAL key is pressed, the measurement mode changes
according to the modes selected in this item.  MANUAL mode is already selected for the mode
selection.