57
When preparation is done, monitoring starts.
NIBP Oscillometric Method
NIBP is measured from the change in amplitude pattern of pulsatile oscillation in cuff pressure
as the cuff pressure is reduced from above systolic to below diastolic pressure.  The occlusive-
oscillometry method uses this to determine the systolic, diastolic and mean arterial pressure.
NIBP Monitoring
Selecting the Initial Cuff Inflation Pressure
The initial cuff inflation pressure can be changed on the PARAMETER SETUP screen.  The
default setting is 180 mmHg.  To change the setting, refer to the “Changing Parameter Setup
Settings” section.
Selecting the Measurement Mode and Interval
Measurement Modes
There are three measurement modes: manual, auto and STAT.  The selected mode or interval is
displayed at the lower right of the screen.
The measurement mode and interval can be changed by pressing the NIBP INTERVAL key.
When the key is pressed, the measurement mode changes according to the modes selected at
“SELECTABLE INTERVALS” on the PARAMETER SETUP screen.  MANUAL mode is
already selected for the mode selection.
To select the modes for the mode selection, refer to the “Changing Parameter Setup Settings”
section.
Manual Measurement
In Manual mode, a single NIBP measurement is performed when the NIBP START/STOP key
is pressed.
STAT (Continuous) Measurement
In STAT mode, measurement is continuously repeated for 15 minutes after the NIBP START/
STOP key is pressed.
When the STAT measurement for 15 minutes is completed, the measurement mode automatically
changes to the Manual mode or Auto mode of selected interval depending on the “NIBP MODE
Monitoring