Operator's Manual  ZM-940PA
59
NOTE
• When measuring patients who are conscious, help the patient to relax.
Measurement may not be accurate if the patient’s arm is tense or if the patient
talks.
• The data for measurement on a leg tends to be higher than measurement on
the arm.  When making diagnosis based on the NIBP values, measure NIBP on
an upper arm.
• Do not apply pressure to the cuff or air hose.  NIBP may not be measured
correctly because of noise or NIBP measurement may stop due to the NIBP
safety circuit.
• When the transmitter is attached to the patient arm and the NIBP
measurement is performed when moving, tell the patient to relax and keep
quiet.  Otherwise, measurement may be stopped or remeasurement is
repeated due to body movement.
• If there is an abnormal noise generated during measurement, stop using the
transmitter and contact your Nihon Kohden representative.
• Do not measure NIBP of a patient on whom an IABP is being used.
Measurement may be incorrect due to the mixing of the patient’s own pulse
and IABP pulse.
• NIBP cannot be measured on a neonate using this transmitter.
1.
Select the measurement mode by pressing the NIBP INTERVAL key.
2.
Press the NIBP START/STOP key to perform measurement.