6.
Every time the NIBP INTERVAL key is pressed, the screen changes as below. Check that
there are no dots missing.
there are no dots missing.
When the NIBP START/STOP key is pressed, the screen returns to the MANUAL
CHECK screen.
CHECK screen.
4. Key Operation
NIBP START/STOP Key
1.
NIBP START/STOP Key
1.
Attach the NIBP cuff to your upper arm.
2.
Press the NIBP START/STOP key. Check that the cuff inflates and deflates properly.
3.
Press the NIBP START/STOP key again. During inflation, press the NIBP START/STOP
key to check that the cuff deflates properly.
key to check that the cuff deflates properly.
NIBP INTERVAL Key
1.
1.
Press the NIBP INTERVAL key and check that the NIBP measuring mode can be changed.
2.
Select any interval and press the NIBP START/STOP key to perform auto measurement.
Check that the NIBP is measured at the selected interval.
Check that the NIBP is measured at the selected interval.