CORITICAL ACTIVITY INDEX 
|
 22
 
 
2.1.3 Frequency Response 
 
Input signal condition 
• 
Amplitude: 200mVpp (Setting waveform generator as 100mVpp) 
• 
Frequency: 2/10/30/40/50(2~48)Hz (EEG), 70/80/100/130/140(80~130Hz)(EMG) 
• 
Waveform:  Sine wave 
 
Measurement method 
• 
Test connection diagram is shown in the above diagram. 
• 
Actual size of signal input is 106uVp. 
• 
Measures the amplitude of detected waveform with a viewer program by changing 
frequency of the waveform generator. 
• 
Obtain benefit (dB unit) per each frequency from the measured amplitude results. 
• 
Benefit = 20*log (numerical benefit) 
 
Measurement results 
Results of brainwave signal channel and electromyogram signal channel measurement are 
shown in below figure. 
(Measurement value at –3 dB passband in EEG, EMG frequency response characteristic test was 
at least 70.7 uVp.) 
 
Figure 2.4 Frequency response curve of brainwave signal channel. -3dB passband: 2~48Hz 
 
 
Figure 2.5 Frequency response curve of electromyogram signal channel. 70~130Hz at -3dB 
frequency band