Charm Engineering Co. Ltd CAI001-BT33 Benutzerhandbuch

Seite von 66
 
CORITICAL ACTIVITY INDEX 
|
 23
 
 
 
 
 
2.1.4 Input Signal Range 
 
Input signal condition 
• 
Amplitude: Increase until output is saturated 
• 
Frequency: 10Hz(EEG), 100Hz(EMG) 
• 
Waveform:  Sine wave 
 
Measurement method 
• 
Connect lines as shown in Fig. 2.1 waveform generator and obtain output signal 
amplitude with viewer program.  
• 
Increase input signal until the output is saturated. Amplitude of output signal at this time 
is maximum input signal amplitude. 
 
Measurement results 
• 
Measurement value: +/-400uVp (EEG); less than +/-550uVp (EMG). 
 
2.1.5 Input Referred Noise 
 
Input signal condition: None 
 
Measurement method 
• 
Connect lines according to below diagram and obtain output signal amplitude with the 
viewer program. At this time, the signal amplitude should be obtained as rms. (rms is 
obtained from 1000 samples) 
 
 
Figure 2.6 Connection diagram to obtain input referred noise size