CORITICAL ACTIVITY INDEX 
|
 39
 
 
 
3.4  Displaying CAI index data 
 
When the sensor inspection is completed successfully, monitoring starts automatically and 
relevant information is displayed on the screen. 
 
 
Figure 3.4 Main screen 
 
3.4.1 Calculation method 
CAI calculation method is to create index based on combination of 4 energy frequencies, offset 
EEG suppression ratio ( BSR, Burst Suppression Ratio), and operate final CAI index. 
 
Calculates brain activity values within the range of 2.5Hz ~ 48Hz  
 
Brain activity is calculated by measuring brainwave value above a certain level by dividing 
measured brainwave into certain time. 
 
Calculates muscular activity by measuring electromyogram in the range of 80Hz ~ 130Hz.