Charm Engineering Co. Ltd CAI001-BT33 Benutzerhandbuch
CORITICAL ACTIVITY INDEX
|
54
3.10 Troubleshooting
Classification Error situation
Error Message
Action to be taken
Sensor
When sensors are not
attached firmly
attached firmly
Sensor Connect
Error
Error
1. Check sensor attachment on the
patient.
2. Check connection between sensor
and amplifier connector.
patient.
2. Check connection between sensor
and amplifier connector.
Signal
When Patient
movement or strong
noise in external
environment occurs
movement or strong
noise in external
environment occurs
Excessive
Artifact
Detected in
Signal
Artifact
Detected in
Signal
1. Remove possible causes of noise
around the sensor and amplifier.
around the sensor and amplifier.
CAI
When the BSR value is
100 for 63s
100 for 63s
Isoelectric EEG
Detected
Detected
1. Check the patient condition.
2. Check whether the sensor is
disconnected.
3. Confirm sensor check.
2. Check whether the sensor is
disconnected.
3. Confirm sensor check.
When the set CAI value
is exceeded
is exceeded
CAI Out of
Target Range
Target Range
1. Check the patient condition.
2. Check the set range.
2. Check the set range.
Battery
When battery remains
of the monitor is below
10%
of the monitor is below
10%
Low Battery
Power
(Monitor)
Power
(Monitor)
1. Recharge the CAI Monitor
When battery remains
of the amplifier is below
10%
of the amplifier is below
10%
Low Battery
Power
(Amplifier)
Power
(Amplifier)
1. Recharge CAIx.