3.10 Troubleshooting 
 
Classification  Error situation 
Error Message 
Action to be taken 
Sensor 
When sensors are not 
attached firmly 
Sensor Connect 
Error 
1. Check sensor attachment on the 
patient. 
2. Check connection between sensor 
and amplifier connector. 
Signal 
When Patient 
movement or strong 
noise in external 
environment occurs 
Excessive 
Artifact 
Detected in 
Signal 
1. Remove possible causes of noise 
around the sensor and amplifier.   
CAI 
When the BSR value is 
100 for 63s 
Isoelectric EEG 
Detected 
1. Check the patient condition. 
2. Check whether the sensor is 
disconnected. 
3. Confirm sensor check. 
When the set CAI value 
is exceeded 
CAI Out of 
Target Range 
1. Check the patient condition. 
2. Check the set range. 
Battery 
When battery remains 
of the monitor is below 
10%  
Low Battery 
Power  
(Monitor) 
1. Recharge the CAI Monitor 
When battery remains 
of the amplifier is below 
10%  
Low Battery 
Power  
(Amplifier) 
1. Recharge CAIx.