CORITICAL ACTIVITY INDEX 
|
 56
 
 
 
4.2  Cleaning and Maintenance of Device 
 
This device should be inspected by personnel approved by the manufacturer at least once a 
year. 
 
4.2.1 Cleaning of device 
When blood or other potentially contaminating material is in contact with this device or its 
parts, users must maintain its cleanness and prevent its malfunction by cleaning prior to usage. 
 
In addition when any liquid or blood is contacted with this device while in use, system 
inspection is also needed to prevent any error in future operation. 
 
* Washing and cleaning of CAI system and its display monitor 
• 
When any liquid or blood is in contact with the CAI system or the monitor, eliminate the 
contaminant immediately using lint-free absorbent towel. (For blood, it may be 
coagulated over time and become very difficult to remove) 
• 
For overall cleaning, wipe with wet towel in detergent, tepid water, and with alcohol 
sequentially and dry it completely.  (Residual moisture may affect inside of the  
connector) 
• 
For disinfection, disinfect it with 10% bleach or bactericide (Lysol(R) professional 
disinfectant foam cleaner spray or PDI disposable wet tissue) on lint-free absorbent towel. 
• 
When cleaning the display screen, wipe it with commercially available display screen 
cleaner or detergent and neutral solution. (However in order to prevent scratch on the 
screen, do not use an abrasive detergent) 
 
* Caution: When the device is used continuously without removal of the blood or liquid, current 
leak may occur. When some disinfectants are mixed up (bleach and ammonia), harmful gas 
may develop.  
 
* Do not use autoclave. Autoclaving may cause damage of product components. 
 
* Avoid liquid ingress to the cable connected with patient. Performance of sensor connector 
may be disturbed.