Charm Engineering Co. Ltd CAI001-BT33 Benutzerhandbuch
CORITICAL ACTIVITY INDEX
|
57
4.2.2 Maintenance
Although CAI system is designed to avoid any need of regular adjustment or calibration,
routine maintenance should be inspected regularly.
* Regular inspection items
•
Check battery remains of CAIx and CAI monitor and contact the manufacturer when
there is any defect in power supply of the battery. (Even though the battery has been
stored after full charging (charging for at least 6 hours), it is necessary to inspect the
battery regularly through operation of CAI monitor.)
there is any defect in power supply of the battery. (Even though the battery has been
stored after full charging (charging for at least 6 hours), it is necessary to inspect the
battery regularly through operation of CAI monitor.)
•
Check whether CAIs and CAIx operate properly after connecting them.
•
As the connection cable is regarded as part of disposables, wearing of the cable is
expected in its continuous usage. Check any contamination, poor contact, damage or
breakage of the cable. (When the connection cable and its connector part have been
used for at least 6 months, perform test on its operation and it is recommended to
replace it when there is any defect.)
expected in its continuous usage. Check any contamination, poor contact, damage or
breakage of the cable. (When the connection cable and its connector part have been
used for at least 6 months, perform test on its operation and it is recommended to
replace it when there is any defect.)
•
Manufactured connection cable has manufacturing date on it to check its aging. (Cable
Part Name-MMDDYY)
Part Name-MMDDYY)
•
Self-Check items for CAI system operation check
Step 1 :
Connect CAIs or Test kit and connection cable of CAIx and then turn CAIx on. When it is
turned on, check its battery remains. (It is recommended to use it when its battery remains
is over 80%).
is over 80%).
Step 2 :
When CAIx and CAI monitor are not turned on after connecting CAIs or Test kit and CAIx
or sensor connection is not identified on the CAI monitor, contact to customer service of
Charm Engineering.
Charm Engineering.
•
When battery of CAIx and CAI monitor has not sufficient power output or is not
continued for at least 1 hour, it is required to replace them with new one.
continued for at least 1 hour, it is required to replace them with new one.