Innokas Yhtyma Oy VC150 Benutzerhandbuch
1-12
VC150 Vital Signs Monitor
KO00065K
Introduction: Product compliance
CAUTIONS
The equipment or system should not be used adjacent to, or stacked
with, other equipment. If adjacent or stacked use is necessary, the
equipment or system should be tested to verify normal operation in the
configuration in which it is being used.
with, other equipment. If adjacent or stacked use is necessary, the
equipment or system should be tested to verify normal operation in the
configuration in which it is being used.
EMC — Magnetic and electrical fields are capable of interfering with the
proper performance of the device. For this reason make sure that all
external devices operated in the vicinity of the monitor comply with the
relevant EMC requirements. X-ray equipment or MRI devices are a
possible source of interference as they may emit higher levels of
electromagnetic radiation. Do not undertake any electrical
modifications to the monitor that are not obtained from Innokas
Medical. Such changes or modifications to the monitor may cause EMC
issues with this or other equipment. This device/system is designed and
tested to comply with applicable standards and regulations regarding
EMC and needs to be installed and put into service according to the
EMC information stated as follows: This device/system is suitable for
use in all establishments other than domestic and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes. Mains power should be
that of a typical commercial or hospital environment.
proper performance of the device. For this reason make sure that all
external devices operated in the vicinity of the monitor comply with the
relevant EMC requirements. X-ray equipment or MRI devices are a
possible source of interference as they may emit higher levels of
electromagnetic radiation. Do not undertake any electrical
modifications to the monitor that are not obtained from Innokas
Medical. Such changes or modifications to the monitor may cause EMC
issues with this or other equipment. This device/system is designed and
tested to comply with applicable standards and regulations regarding
EMC and needs to be installed and put into service according to the
EMC information stated as follows: This device/system is suitable for
use in all establishments other than domestic and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes. Mains power should be
that of a typical commercial or hospital environment.
NOTE
Medical electrical equipment requires special electromagnetic compatibility
(EMC) precautions which must be considered when installing and putting
this equipment into operation. Refer to the service manual for information.
(EMC) precautions which must be considered when installing and putting
this equipment into operation. Refer to the service manual for information.
Exergen temporal scanner
The Exergen temporal scanner has these additional classifications:
•
•
Type BF applied part
•
Internally powered (battery operated)
•
IPX0, degree of protection against ingress of water