1-12
VC150 Vital Signs Monitor
KO00065K
Introduction: Product compliance
CAUTIONS
The equipment or system should not be used adjacent to, or stacked 
with, other equipment. If adjacent or stacked use is necessary, the 
equipment or system should be tested to verify normal operation in the 
configuration in which it is being used.
EMC — Magnetic and electrical fields are capable of interfering with the 
proper performance of the device. For this reason make sure that all 
external devices operated in the vicinity of the monitor comply with the 
relevant EMC requirements. X-ray equipment or MRI devices are a 
possible source of interference as they may emit higher levels of 
electromagnetic radiation. Do not undertake any electrical 
modifications to the monitor that are not obtained from Innokas 
Medical. Such changes or modifications to the monitor may cause EMC 
issues with this or other equipment. This device/system is designed and 
tested to comply with applicable standards and regulations regarding 
EMC and needs to be installed and put into service according to the 
EMC information stated as follows: This device/system is suitable for 
use in all establishments other than domestic and those directly 
connected to the public low-voltage power supply network that 
supplies buildings used for domestic purposes. Mains power should be 
that of a typical commercial or hospital environment.
NOTE
Medical electrical equipment requires special electromagnetic compatibility 
(EMC) precautions which must be considered when installing and putting 
this equipment into operation. Refer to the service manual for information.
Exergen temporal scanner
The Exergen temporal scanner has these additional classifications:
Type BF applied part
Internally powered (battery operated)
IPX0, degree of protection against ingress of water