ABB WICB Benutzerhandbuch

Seite von 72
24   OI/266/WIHART-EN | 2600T Series Pressure transmitters
5 Mounting
5.13.6 Pressure or absolute pressure measurement of a 
tank
  — Place the taps in the upper part of the tank.
  — Mount the transmitter above the elevation of the process    
    tap (both pressure and differential pressure transmitter can  
    be used).
 — 
Connect the transmitter to the tank.
Figure 27: Gauge or absolute pressure measurement on a tank
5.13.7 Pressure or absolute pressure measurement of a 
liquid in a pipe
  — Place the tap at the side of the line.
  — Mount the transmitter (both pressure and differential  
 
    pressure transmitters) beside or below the tap for clean    
    fluids, above the tap for dirty fluids.
  — Connect the + (H) side of the transmitter to the pipe.
Figure 28: Gauge or absolute pressure measurement of a liquid in 
a pipe