58 OI/266/WIHART-EN | 2600T Series Pressure transmitters
9 Troubleshooting
9.4 HMI as diagnostic indicator
While the HMI works as Variable Indicator, also diagnostic
strings can be displayed. Whenever a failure or warning
condition is detected within the transmitter, a message appears
in the low side of the display below the bar-graph. The
message is formed by the NAMUR NE107 icon and the string
of the component where the problem occurred.
strings can be displayed. Whenever a failure or warning
condition is detected within the transmitter, a message appears
in the low side of the display below the bar-graph. The
message is formed by the NAMUR NE107 icon and the string
of the component where the problem occurred.
NAMUR
icons
icons
Description
Error / Failure
Functional check (e.g. during simulation )
Out of Spec (e.g. Sensor temperature outside the specs limits)
Maintenance required
Figure 42: Example of “Maintenance / Sensor” and “Failure /
Electronics” diagnostic
Electronics” diagnostic
Detailed diagnostic info from HMI
When the above kind of diagnostic information is displayed,
from the HMI it is also possible to see the details.
from the HMI it is also possible to see the details.
1.
Press the center navigation keys for 4 seconds until two
special icons appear on the bottom corners of the display
Important. This step is necessary only for the keypad activation of
TTG HMI type. For conventional HMI start from step 2.
TTG HMI type. For conventional HMI start from step 2.
2.
Press the left navigation key
3.
The HMI enter in the special menu with three items:
Diagnostics
Operator View 1
Signals View
4.
Select “Diagnostics” and the list of all the active error
conditions appears with on top the worst condition.
The format of how the detailed diagnostic info are displayed is
“XA.BBB” where:
“XA.BBB” where:
X = NAMUR NE107
Categories
Categories
A = Priority.
(Higher number =
higher priority)
(Higher number =
higher priority)
BBB = error
code
code
F = Failure
M = Maintenance
O = Out of Specification
C = Function Check
M = Maintenance
O = Out of Specification
C = Function Check
9.5 HMI as feedback of the local push button
operations
As consequence of the operations described in the section 8.2,
when the Z or S buttons are released, the feedback of the
executed operation is displayed in the bottom of the LCD
(same position as per diagnostic messages):
when the Z or S buttons are released, the feedback of the
executed operation is displayed in the bottom of the LCD
(same position as per diagnostic messages):
Message
Description
! Oper Done
The push button operation has been successfully executed
! Proc Too
Low
Low
The Pressure measured in input is too low and not
acceptable for the requested operation
acceptable for the requested operation
! Proc Too
High
High
The Pressure measured in input is too high and not
acceptable for the requested operation
acceptable for the requested operation
! New URV
Error
Error
The Zero (Z) operation cannot be accepted because the
URV would be shifted outside the Upper Sensor limit
URV would be shifted outside the Upper Sensor limit
! Span Error
The Span (S) operation cannot be accepted because the
new URV would be too close to the LRV and their
difference lower than the Minimum Span value
new URV would be too close to the LRV and their
difference lower than the Minimum Span value
! Oper
Disabled
Disabled
The push button operation has been refused because the
Write Protection is enabled.
Write Protection is enabled.
! LRV Too
Low
Low
New LRV is too low and not acceptable for the requested
operation
operation
LRV Too
High
High
New LRV is high low and not acceptable for the requested
operation
operation
URV Too
Low
Low
New URV is too low and not acceptable for the requested
operation
operation
URV Too
High
High
New URV is high low and not acceptable for the requested
operation
operation
Armed
Device is armed to accept HART command 73 “Find
Device”. This message can be triggered only during the
device wakeup operation
Device”. This message can be triggered only during the
device wakeup operation
Possible sources of error are: ELECTRONICS, SENSOR,
PROCESS and CONFIGURATION. These sources can show up
the above listed icons according to the specific case.
PROCESS and CONFIGURATION. These sources can show up
the above listed icons according to the specific case.