Olympus Winter & Ibe GmbH ESG200 Benutzerhandbuch
79
Troubleshooting
Problem
Cause
Remedy
The electrosurgical
generator does
not respond to foot
switch or hand switch
activation.
Improper contact of the connection
cable of the foot switch or the HF
instrument to the corresponding
socket of the electrosurgical
generator.
Check the connection cable of the
foot switch and the HF instrument for
correct connection.
The foot switch, the hand switch of
the HF instrument or their connection
cables are damaged.
Observe the foot switch, the hand
switch of the HF instrument and their
connection cables for damages. If
necessary, replace the foot switch, the
HF instrument or the corresponding
connection cable.
The wrong foot switch pedal or the
wrong hand switch is pressed.
Press the correct foot switch pedal
or the correct hand switch of the HF
instrument.
The electrosurgical generator is not
switched on.
Switch on the electrosurgical
generator with the power switch.
2 switches for activating the output
power are pressed simultaneously,
e.g. another foot switch pedal or hand
switch of an HF instrument.
Release all switches for activation
and check that no switch is pressed
unintentionally. Then activate the
output power using only 1 switch.
A message box is displayed on the
touch screen.
Tap [OK] or [Cancel] to close the
message box or wait until the
message box closes automatically
after a few seconds.
It is possible to activate the output
power only if either the All Screen or
the Set Screen is displayed on the
touch screen.
Return to the All Screen or the Set
Screen.
The output mode is deactivated ([Off]
is displayed) or the output power level
is set to 0 ([--] is displayed).
Activate the output mode on the Mode
Screen or increase the output power
level on the Set Screen. Refer to the
section “Selecting the appropriate
output settings” on page 62.
Malfunction of the electrosurgical
generator.
Do not use this electrosurgical
generator.
Contact Olympus.
Autostart is selected
and the HF instrument
has contact with
the tissue, but the
electrosurgical
generator does not
activate.
A long time delay for Autostart is set. Reduce the time delay for Autostart.
Refer to the section “Autostart Setup”
on page 45.
Malfunction of the electrosurgical
generator.
Do not use this electrosurgical
generator.
Contact Olympus.
DRAFT
–
2017-01-17