Troubleshooting
Problem
Cause
Remedy
The electrosurgical 
generator does 
not respond to foot 
switch or hand switch 
activation.
Improper contact of the connection 
cable of the foot switch or the HF 
instrument to the corresponding 
socket of the electrosurgical 
generator.
Check the connection cable of the 
foot switch and the HF instrument for 
correct connection.
The foot switch, the hand switch of 
the HF instrument or their connection 
cables are damaged.
Observe the foot switch, the hand 
switch of the HF instrument and their 
connection cables for damages. If 
necessary, replace the foot switch, the 
HF instrument or the corresponding 
connection cable.
The wrong foot switch pedal or the 
wrong hand switch is pressed.
Press the correct foot switch pedal 
or the correct hand switch of the HF 
instrument.
The electrosurgical generator is not 
switched on.
Switch on the electrosurgical 
generator with the power switch.
2 switches for activating the output 
power are pressed simultaneously, 
e.g. another foot switch pedal or hand 
switch of an HF instrument.
Release all switches for activation 
and check that no switch is pressed 
unintentionally. Then activate the 
output power using only 1 switch.
A message box is displayed on the 
touch screen.
Tap [OK] or [Cancel] to close the 
message box or wait until the 
message box closes automatically 
after a few seconds.
It is possible to activate the output 
power only if either the All Screen or 
the Set Screen is displayed on the 
touch screen.
Return to the All Screen or the Set 
Screen.
The output mode is deactivated ([Off] 
is displayed) or the output power level 
is set to 0 ([--] is displayed).
Activate the output mode on the Mode 
Screen or increase the output power 
level on the Set Screen. Refer to the 
section “Selecting the appropriate 
output settings” on page 62.
Malfunction of the electrosurgical 
generator.
Do not use this electrosurgical 
generator.
Contact Olympus.
Autostart is selected 
and the HF instrument 
has contact with 
the tissue, but the 
electrosurgical 
generator does not 
activate.
A long time delay for Autostart is set.  Reduce the time delay for Autostart. 
Refer to the section “Autostart Setup” 
on page 45.
Malfunction of the electrosurgical 
generator.
Do not use this electrosurgical 
generator.
Contact Olympus.