6-19
Measurement Setting
1. Source Selection
Press Measure Source to select the desired channel for measurement (CH1,
CH2 or MATH). The color of the parameter icons under MENU at the left of the
screen will change with the source selected.
CH2 or MATH). The color of the parameter icons under MENU at the left of the
screen will change with the source selected.
2. Measurement Range
Press Measure Meas.Range Meas.Range to select “Screen Region” or
“Cursor Region” for measurement. When “Cursor Region” is selected, two cursor
lines appear on the screen. At this point, press CursorA and CursorB and use
“Cursor Region” for measurement. When “Cursor Region” is selected, two cursor
lines appear on the screen. At this point, press CursorA and CursorB and use
to adjust the position of the two cursor lines respectively so as to determine
the measurement range. Or, press CursorAB and use
to adjust the
positions of cursor A and cursor B at the same time. Note that you can press
continuously to switch between the cursors.
3. Delayed Measurement Setting
Specify the source A and source B in the measurement items “Delay A→B ”
and “Delay A→
and “Delay A→
B ”.
Press Measure Meas.Setting Type “Delay” and then press SourceA
and SourceB to set the two channel sources (CH1 or CH2) of delayed
measurement respectively.
and SourceB to set the two channel sources (CH1 or CH2) of delayed
measurement respectively.