  DS2000A User’s Guide 
6-20 
4.  Phase Measurement Setting 
Specify the source A and source B in the measurement items “Phase A” 
and “Phase A
”. 
Press Measure  Meas.Setting  Type  “Phase” and then press SourceA 
and SourceB to set the two channel sources (CH1 or CH2) of phase 
measurement respectively.   
 
5.  Threshold Measurement Setting 
Specify the vertical level (in percentage) being measured in the analog channel. 
Measurements of all the time, delay and phase parameters will be influenced by 
this setting.   
Press Measure  Meas.Setting  Type  “Threshold” and then: 
  Press Max and use 
  to set the maximum value of the measurement. 
Reducing the maximum value to the current “Mid” will automatically reduce 
the “Mid” to keep it lower than the maximum value. The default is 90% and 
the range available is from 7% to 95%. 
  Press Mid and use 
  to set the middle value of measurement. The 
middle value is limited by the settings of “Max” and “Min”. The default is 50% 
and the range available is from 6% to 94%. 
  Press Min and use 
  to set the minimum value of the measurement. 
Increasing the minimum value to the current “Mid” will automatically 
increase the “Mid” to keep it higher than the minimum value. The default is 
10% and the range available is from 5% to 93%.