12. Set the external trigger level using 
the side menu.  
 
 
 
Range 
00.0V~ 5 screen divisions 
 
 
 
Set to TTL 1.4V 
 
 
 
Set to ECL -1.3V 
 
 
 
Set to 50% 
Using Advanced Delay Trigger  
 
Background 
The EXT trigger source is always used as the delay 
triggering source.  
Panel Operation 
1.  Press the trigger Menu key. 
Menu
 
 
2.  Press Type from the lower bezel 
menu.   
 
 
3.  Select Delay from the side menu. 
The delay + edge trigger indicator 
appears at the bottom of the 
display. 
 
 
 
 
 
From left: delay trigger, delay source 
(external), delay coupling + edge source, edge 
slope, edge trigger level 
 
4.  To set the delay press Delay from 
the bottom bezel.