CONFIGURATION 
155 
 
5.  To Delay by Time, press Time from 
the side menu and set the delay 
time.  
 
 
 
Range 
10ns ~ 10s (by time) 
Set to minimum 
 
6.  To Delay by Event, press Event 
from the side menu and set the 
number of events.  
 
 
 
Range 
1 ~ 65535  events 
 
 
 
Set to Minimum 
Using Pulse Width Trigger 
 
Panel Operation 
1.  Press the trigger Menu key. 
Menu
 
 
2.  Press the Type key from the lower 
bezel menu.  
 
 
3.  Select Pulse Width from the side 
menu. The pulse width trigger 
indicator appears at the bottom of 
the display. 
 
 
 
 
From left: source, polarity, when, coupling 
 
4.  Press Source from the lower bezel.