Lecroy WAVEACE 1001 2-channel oscilloscope, Digital Storage oscilloscope, WaveAce 1001 Benutzerhandbuch

Produktcode
WaveAce 1001
Seite von 89
Operator's Manual 
920836 Rev A  
57 
STATUS - CMR? 
Description 
The CMR? query reads and clears the contents of the CoMmand error Register (refer to the 
following table for details) which specifies the last syntax error type detected by your 
oscilloscope. 
Query Syntax 
CMR? 
Response Format 
CMR <value> 
<value>:= 0 to 13 
Related Commands 
ALL_STATUS?, *CLS 
CMR? Status Register Structure 
Value  Description 
Unrecognized command/query header. 
Illegal header path. 
Illegal number. 
Illegal number suffix. 
Unrecognized keyword. 
String error. 
GET embedded in another message. 
10 
Arbitrary data block expected. 
11 
Non-digit character in byte count field of arbitrary data block. 
12 
EOI detected during definite length data block transfer. 
13 
Extra bytes detected during definite length data block transfer.