57
STATUS - CMR?
Description
The CMR? query reads and clears the contents of the CoMmand error Register (refer to the
following table for details) which specifies the last syntax error type detected by your
oscilloscope.
following table for details) which specifies the last syntax error type detected by your
oscilloscope.
Query Syntax
CMR?
Response Format
CMR <value>
<value>:= 0 to 13
Related Commands
ALL_STATUS?, *CLS
CMR? Status Register Structure
Value Description
1
Unrecognized command/query header.
2
Illegal header path.
3
Illegal number.
4
Illegal number suffix.
5
Unrecognized keyword.
6
String error.
7
GET embedded in another message.
10
Arbitrary data block expected.
11
Non-digit character in byte count field of arbitrary data block.
12
EOI detected during definite length data block transfer.
13