Source 1 (DATA) Level Type and Vertical Level - The message decoding algorithm setup is performed 
here. The level is normally set up in %, and defaults to 50%. Adjust the level by touching inside the field 
and highlight the box title, then use the oscilloscope front panel Adjust knob to make the change. 
Alternatively, touch inside the field twice and select a value using the pop-up numeric keypad. 
Creating a UART-RS232bus Trigger Condition 
The UARTbus Trigger dialog, with detail on some of the setup conditions, is shown in the following topics. 
PLEASE NOTE THE FOLLOWING:  
 
The RS-232 Trigger dialog is nearly the same, but contains less flexibility. Therefore, only the UART Trigger 
dialog is described here. 
 
Refer to Using The D Supported Protocol Toolsets (on page 14) to correctly access the Trigger Condition 
dialog specific to your desired protocol. 
 
The Source and UARTbus Setup information must be defined. The datasheet for your part should contain the 
information you need to properly setup the UART Trigger.  
Selection of Trigger Type results in dynamic changes to the UART Trigger dialog. Simple Parity ERROR triggering 
requires no additional setup, while DATA triggers require defining of the Data Pattern, selection of Condition, 
etc. Also, if you are looking for the exact Position of DATA, then the Interframe Time must be defined.  
Select condition values by touching fields (using your finger, or use a mouse pointer). A pop-up is shown where 
you can choose from EqualNot EqualLess thanLess than or Equal toGreater thanGreater than or Equal to
In Range, or Out Range condition values.