Lecroy WSXS-I2CBUS TD WSXS-I2CBUS TD trigger and decoder extension Compatible with LeCroy WaveSurfer® oscilloscopes. WSXS-I2CBUS TD Benutzerhandbuch

Produktcode
WSXS-I2CBUS TD
Seite von 204
Serial Data Debug Solutions 
66 
919586 RevA 
UART-RS232bus Trigger Setup Detail 
The following topic provides specific control settings for a UARTbus Trigger (again, since they're almost identical, 
but UART contains a bit more detail than RS-232). 
 
The previously numbered UARTbus trigger sections correspond with the following explanations. 
1.
 
S
OURCES 
S
ETUP
 
 
DATA - The pop-up dialog is used to select the appropriate channel or EXT inputs for each. Set these fields 
up with caution or your trigger may not function correctly. 
 
Threshold (Trigger) - Adjust the vertical level for the trigger. Much like an Edge trigger, you must specify 
the level used to process the incoming signals and determine whether the desired serial data pattern is 
meeting the set trigger condition. 
2.
 
UART
 
S
ETUP
 
 
Bitrate - Use the Bitrate field to adjust the value and match the bus to which you are connected. This 
bitrate selection is dynamically linked to the decoding bitrate (they are always the same value). Use the 
arrows to move through standard bit rates (300 b/s, 1.2, 2.4, 4.8, 9.6, 19.2, 28.8, 38.4, 57.6, 76.8, 115.2, 
230.4, 460.8, 921.6, kb/s, 1.3824 1.8432, 2.7648 Mb/s) and make a selection. Or, touch the number twice 
(with a finger, or using a mouse) and open a pop-up keypad and enter the value directly, anywhere 
between 30 b/s and 500 Mb/s. 
 
Data Bits - Select the number of data bits per byte (not including the START, STOP, or PARITY bits). Trigger 
on UART with a 9
th
 DATA bit used as an Alert bit by entering Data Bits = 9, and then define the 9
th
 Alert bit 
as a 0, 1, or X (don’t care) as needed. 
 
Parity - Choose from OddEven, or None in the Parity field. Only when Odd or Even values are made in 
this field is the Parity Error Trigger Type enabled. 
 
Stop Bits - Choose 11.5, or 2 Stop Bits in this field. 
 
Bit Order - Choose either Most Significant Bit (MSB) or Least Significant Bit (LSB) bit order in this field. 
Note: This field defaults to LSB and cannot be changed on an RS-232 trigger. 
 
Polarity - Choose the Polarity of the UART signal as either IdleLow (Data 1 = High) or IdleHigh (Data 1 = 
Low). 
Note: This field defaults to IdleLow and cannot be changed on an RS-232 trigger. 
3.
 
T
RIGGER 
T
YPE
 
The Data button is selected by default unless Odd or Even Parity is selected on the Parity field. Then, the 
Parity Error Trigger Type button is enabled for use. 
4.
 
S
ETUP 
F
ORMAT
 
Select either Binary or Hexadecimal (Hex) setup mode. The mode selected affects the format of the 
following Data Value and Data Value To fields.