OPERATION 
45 
PASS / FAIL Test Judgments 
 
Background 
If the test is allowed to run to completion (the 
test is not stopped or a protection setting is not 
tripped) then the tester will judge the test as 
either PASS or FAIL.  
 Note 
The test will be judged PASS when: 
 
The H and L limits (high and low limits) 
have not been tripped during the test time. 
The test will be judged FAIL when: 
 
Either the H or L limit has been tripped 
during the test time. 
 
The ARC current limit has been tripped 
during a test (if enabled).  
PASS Judgment 
1.  When the test is judged as PASS, 
PASS will be displayed, the buzzer 
will sound once and the PASS 
indicator will be lit green.  
PASS
 
 
 
     
 
 
 
H = 2 2 . 0 m A  
 
 
 
 
m A
L
0
=
0 0
.
   
6 0 S  
 
 
 
 
 
= 0
T
       
   
CW
D
A CW
R  
I
D
O E
M
T M E
I
1 00
 
 
0 90
 
k V
S
A
P
 
S
A
m
 
 
 
 
2.  The PASS judgment will be held on the display 
until the STOP or START button is pressed. 
Pressing the STOP button will 
return the tester to the READY 
status. 
STOP