OPERATION 
47 
FAIL Judgment 
1.  When the test is judged as FAIL, 
FAIL will be displayed, the buzzer 
will sound and the FAIL indicator 
will be lit red. 
As soon as a test is judged FAIL, 
the output is cut from the 
terminals. 
FAIL
 
 
 
     
 
 
 
H = 2 2 . 0 m A  
 
 
 
 
m A
L
0
=
0 0
.
   
6 0 S  
 
 
 
 
 
= 0
T
       
   
CW
D
A CW
R  
I
D
O E
M
T M E
I
 
1 00
 
k V
I
A
F
 
L
A
m
 
23 0
 
 
 
 
2.  The FAIL judgment will be held on 
the display until the STOP button 
is pressed twice. Pressing the STOP 
button once will stop the buzzer, 
pressing the STOP button again 
will return the tester to the READY 
status.   
STOP
x2 
 
 
     
 
 
 
H = 2 2 . 0 m A  
 
 
 
 
m A
L
0
=
0 0
.
   
7 4 S  
 
 
 
 
 
= 1
T
       
   
CW
D
A CW
R  
I
D
O E
M
T M E
I
 
1 00
 
k V
A
E
R
 
D
 
Y
A
m
 
FAIL Timing 
Diagrams 
The timing diagrams below show the ACW, DCW 
and IR timing for the START status, TEST status 
and FAIL judgment. 
ACW FAIL Timing 
Output V
time
START
TEST
FAIL
RAMP
Discharge time 
TEST TIME
(100ms)