Gw Instek GPT-9801 Insulation measuring device, GPT-9801 Benutzerhandbuch

Produktcode
GPT-9801
Seite von 156
 
 
GPT-9000 Series User Manual 
120
 
MANU:STEP 
Set
Query
 
 
Description 
Sets the MANU test number.    
Syntax 
Query Syntax 
MANU:STEP <NR1> 
MANU:STEP? 
Parameter/ 
Return parameter 
<NR1> 
0~100. 
Example 
MANU:STEP 100 
Sets the manual test number to 100. 
MANU:NAME 
Set
Query
 
 
Description 
Sets or returns the test name for the selected 
manual test. The test must be in MANU mode 
before this command can be used.   
Note only alphanumeric characters (A-Z, a-z, 0-9) 
and the “_” underscore character can be used to set 
the MANU test name.  
Syntax 
Query Syntax 
MANU:NAME <string> 
MANU:NAME? 
Parameter/ 
Return parameter 
<string> 
10 character string. (first character must 
be a letter) 
Example 
MANU:NAME test1 
Sets the manual test name to “test1”. 
MANU:RTIMe 
Set
Query
 
 
Description 
Sets or returns the Ramp Time for the test in 
seconds.  
 
Note: A “TIME ERR” will result  if the Ramp Time 
+ Test Time is ≥ 240 seconds when the HI SET limit 
is over 30mA (ACW function only).