Gw Instek GPT-9801 Insulation measuring device, GPT-9801 Benutzerhandbuch

Produktcode
GPT-9801
Seite von 156
 
REMOTE CONTROL 
121 
Syntax 
Query Syntax 
MANU:RTIMe <NR2> 
MANU:RTIMe? 
Parameter/ 
Return parameter 
<NR2> 
0.1~999.9 seconds  
Example 
MANU:RTIM 0.5 
Sets the ramp time to half a second. 
MANU:EDIT:MODE 
Set
Query
 
 
Description 
Sets or returns the mode (ACW, DCW, IR) of the 
selected manual test.  
Syntax 
Query Syntax 
MANU:EDIT:MODE {ACW|DCW|IR|GB} 
MANU:EDIT:MODE? 
Parameter/ 
Return parameter 
<ACW> 
AC Withstand mode 
<DCW> 
DC Withstand mode 
<IR> 
Insulation Resistance mode 
Example 
MANU:EDIT:MODE ACW 
Sets the mode to ACW. 
MANU:ACW:VOLTage 
Set
Query
 
 
Description 
Sets or returns the ACW voltage in kV.  The test 
must first be in ACW mode before this command 
can be used.  
Syntax 
Query Syntax 
MANU:ACW:VOLTage <NR2> 
MANU:ACW:VOLTage? 
Parameter/ 
Return parameter 
<NR2> 
0.100 ~ 5.000 (kV) 
Example 
MANU:ACW:VOLT 1 
Sets the ACW voltage to 1 kV.