Gw Instek GPT-9801 Insulation measuring device, GPT-9801 Benutzerhandbuch

Produktcode
GPT-9801
Seite von 156
 
 
GPT-9000 Series User Manual 
62
 
PASS / FAIL MANU Test 
 
Background 
If the test is allowed to run to completion (the 
test is not stopped or a protection setting is not 
tripped) then the tester will judge the test as 
either PASS or FAIL.  
 Note 
The test will be judged PASS when: 
 
The HI SET and LO SET limits have not been 
tripped during the test time. 
The test will be judged FAIL when: 
 
Either the HI SET or LO SET limit has been 
tripped during the test time. 
 
A protection setting has been tripped during 
the test time. See page 148 for a list of error 
messages. 
PASS Judgment 
1.  When the test is judged as PASS, 
PASS will be displayed, the buzzer 
will sound and the PASS indicator 
will be lit green.  
PASS
 
 
I R  
G B  
 
A CW
D CW
 
 
E        
 
 
F R E Q =  
0 H z  
6
0
   
100
k
 
V
 
       
R
A C
 
=
F F
O
S
A
P
 
                         
M
T I
E
= 0 0 1 . 0 S
R
0 0 m A
E
R F
=
#
0 .
0
1 . 0 0 m A
I
H
E
S
= 0
T
0 0 . 1 S
0
A M P
=
R
 
M
M A U
N
N A
_
2
M A N U = * * * - 0 0
 
 
S
m A
00 37
 
 
 
2.  The PASS judgment will be held on the display 
until the STOP or START button is pressed. 
Pressing the STOP button will 
return the tester to the READY 
status. 
STOP