Gw Instek GPT-9801 Insulation measuring device, GPT-9801 Benutzerhandbuch

Produktcode
GPT-9801
Seite von 156
 
OPERATION 
63 
Pressing the START button will 
restart the test. 
START
 
 Note 
The buzzer will only sound if the Pass Sound is set 
to ON. See page 73 for details. 
 
The START button is disabled when the buzzer is 
beeping. 
PASS Timing 
Diagrams 
The timing diagrams below show the ACW, DCW, 
IR and GB timing for the START status, TEST 
status and PASS judgment. 
ACW PASS 
Timing  
Discharge time 
Output V
time
Initial time 
(100ms)
START
TEST
PASS
RAMP
TEST TIME
 
DCW PASS 
Timing 
Discharge time 
Output V
time
Initial time 
(100ms)
START
TEST
PASS
RAMP
TEST TIME