GPT-9000 Series User Manual
68
4. The ZERO function can be
activated by pressing the
corresponding soft-key in the
READY status. The ZERO soft-key
will be highlighted.
corresponding soft-key in the
READY status. The ZERO soft-key
will be highlighted.
E R O
Z
5. Press the START button to perform
the zero check. The tester will go
into the ZERO status.
into the ZERO status.
START
0
3 01
A
Z E R 0
I R
G B
A CW
D CW
E
F R E Q =
0 H z
6
R
E
Z
G B V = 0 . 6 2 2 V
M
T I
E
= 0 0 3 . 2 S
R
. 0 m Ω
E
R F
=
#
0 0
0
0 0 . 0 m Ω
I
H
E
S
= 3
T
M
M A U
N
N A
_
2
M A N U = * * * - 0 0
O
Measured Resistance
200 7
m Ω
ZERO status
6. When the zero check has finished, the tester
will return back to the VIEW status. The
resistance of the test leads will be automatically
set as the Reference value.
resistance of the test leads will be automatically
set as the Reference value.
0
3 01
A
I R
G B
A CW
D CW
E
F R E Q =
0 H z
6
E
I
V
G B V = V
M
T I
E
= 0 0 3 . 2 S
R
. 7 m Ω
E
R F
=
#
0 0
2
0 0 . 0 m Ω
I
H
E
S
= 3
T
M
M A U
N
N A
_
2
M A N U = * * * - 0 0
W
m Ω
Reference value
Note
Remember to replace the test leads to the proper
position on the DUT before testing.
position on the DUT before testing.