GPT-9000 Series User Manual 
70
 
Special MANU Test Mode (000) 
 
Background 
When MANU number 000 is selected, the 
special test mode is activated. Under the special 
test mode, the voltage can be changed during a 
test, in real time (ACW, DCW only). The test 
function can also be changed when in READY 
or VIEW status, unlike under normal operation.  
Separate settings can be saved under the special 
test mode for each of the testing functions: 
ACW, DCW, IR and GB. This means a different 
ACW, DCW, IR and GB test setup can be saved 
for MANU number 000. 
Steps 
1.  Choose MANU number 000 to 
enter the special test mode. 
 
2.  Set all the necessary parameters for 
a test and save. A different test 
setup can be saved for each test 
function (ACW, DCW, IR and GB). 
 
 
I R  
G B  
A CW
D CW
 
 
E        
 
 
F R E Q =  
0 H z  
6
0
   
100
k
 
V
 
       
R
A C
 
=
F F
O
E
I
V
 
 
                         
M
T I
E
= 0 0 1 . 0 S
R
0 0 m A
E
R F
=
#
0 .
0
1 . 0 0 m A
I
H
E
S
= 0
T
0 0 . 1 S
0
A M P
=
R
 
W
M
M A U
N
N A
_
0
M A N U = * * * - 0 0
 
 
m A
 
 Note 
The TIMER settings can be set OFF when in the 
special test mode for ACW and DCW tests. 
 
3.  The settings of a different test 
function can be loaded by pressing 
the corresponding soft-key in the 
VIEW or READY status. 
A CW
 
Example: 
ACW