Gossen Metrawatt Metrahit Iso Insulation measuring device, CAT III 600 V Metrahit Iso Benutzerhandbuch

Produktcode
Metrahit Iso
Seite von 72
GMC-I Gossen-Metrawatt GmbH
29
Measurements: V/Hz, 
Ω,  , Temp (°C/°F), 
, M
Ω
@UISO
, A/Hz
Specified measuring accuracy is not reached with signals of 
greater than 500 Hz when the filter is active.
Voltage Comparator for Displaying Dangerous Voltage
The input signal or measuring signal is checked by a voltage 
comparator for dangerous spikes, because these do not appear 
at the display when the low-pass filter is used. 
At voltages of greater than 15 V AC or 25 V DC, a danger symbol 
appears at the display: 
.
COM mA A
V
Temp
Ω MΩ
ISO
Max. 1000 V, 3 kHz
V~: 10 mV 
… 1000 V
Hz: 1 Hz 
… 300 kHz
~
P
max
 = 3 x 10
6
 V x Hz
Measuring Ranges:
> 15 V AC 
or > 25 V DC: 
> 1000 V:
Warnings regarding dangerous voltage:
230
.
0
5 ranges: 300 mV, 3 V, 30 V,
300 V and 1000 V
0
230.6
V
AC
TRMS
0
050.3
Hz
AC
V~
Hz / %
V~
Hz
0
229.9
V
AC
TRMS
Fil
V~ with Filter
1 kHz
V~
0
050.0
AC
Fil
Hz
Hz with Filter
!
!
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
1 kHz
Hz
MEASURE
SETUP
 ... 
Z
 
SET 
 ... 
V
 
CL IP
FUNC
ENTER
FUNC
ENTER
  
UV
 
0ff
%
%
Hz: 
vpm 
= 0FF
RPM: 
vpm 
≠ 0FF
The following function is 
only available with cus-
tomer-specific variants.