Measurements: V/Hz, 
Ω,  , Temp (°C/°F), 
, M
Ω
@UISO
, A/Hz
RPM Measurement (function only available with customer-specific variant)
Revolutions per minute (also known as rotational frequency) are 
measured by acquiring pulses. As a prerequisite for this measure-
ment, the number of measurable pulses per revolution must first 
be set in the RPM setup menu (
vpm 
≠ 0FF, see below).
➭ Set the rotary switch to V~. 
➭ Repeatedly press the FUNC | ENTER multifunction key until 
RPM is briefly displayed. The measured value then appears in 
RPM, for example “244.3 r”. 
Measured value RPM = revolutions per minute.
Parameter 
vpm = pulses per revolution.
Pulses per Revolution Setup Menu
 
1nFo 
Z
 ... 
Z
 
SET 
 rAtE 
V
 ... 
V
 
vpm 
 
0ff/1 :1 ... 1000 
Y
 
ZUV
 
5.1.3 Transient Overvoltages
The multimeters are protected against transient overvoltages of up 
to 6 kV with wave-front durations of 1.2 ms and halftimes of 50 
μs 
in the voltage measuring range. If longer pulse durations are ex-
pected, e.g. when conducting measurements at transformers or 
motors, use of our KS30 measuring adapter is recommended. It 
provides protection against transient overvoltages of up to 6 kV 
with wave-front durations of 10, and halftimes of 1000
μs. Continu-
ous load capacity is 1200 V
RMS
. Additional influence error caused 
by the KS30 measuring adapter amounts to approximately –2%.
5.1.4 Voltage Measurements at Above 1000 V
Voltages of greater than 1000 V can be measured with a 
high-voltage probe, e.g. the HV3
1
 or the HV30
2
 from 
GMC-I Gossen-Metrawatt GmbH. It is absolutely essential to 
earth the ground terminal in this case. Observe all applicable 
safety precautions!
1
HV3: 3 kV
2
HV30: 30 kV, for   (DC) voltages only
RPM
Revolutions
min
--------------------------------
Pulses
Revolution
-----------------------------
⎞ x60s
s
--------
=